Chinese Journal of Applied Chemistry ›› 2009, Vol. 26 ›› Issue (12): 1495-1497.

• Original Articles • Previous Articles     Next Articles

Analysis of Opiates in Human Blood by Solid-Phase Extraction and Gas Chromatography/ Mass Spectrometry Detection

WANG Yan-Yan,MENG Pin-Jia*   

  1. (Department of Forensic Science,Chinese People′s Public Security University,Beijing 100038)
  • Received:2008-11-27 Revised:2009-05-06 Published:2009-12-10 Online:2009-12-10

Abstract:

Morphine, codeine and 6-acetyl morphine in human blood was detected using solid-phase extraction and gas chromatography/mass spectrometry-selected ion monitoring, with ethyl morphine as internal standard, 3 opiates spiked in blank blood were extracted using a solid-phase cartridge. The fluent was evaporated to dryness followed by being derivatized with MBTFA, then it was analyzed directly by GC/MS in selected ion mode. The method was validated in the range 10~1000ng/mL with the drug-spiked blood, the reactor responded linearly over the studied range (R2>0.99 for the 3 drugs); The relative standard deviations (RSD) for intra and inter-days were less than10% and 16%, respectively the relative extraction recoveries vary from 79% to 116%, and LODs were less than 1, 0.5 and 3ng/mL, which promised the perspective application of the method in detecting trace amount of opium in human blood.

Key words: opiate,GC/MS-SIM detection,solid-phase extraction

CLC Number: