Chinese Journal of Applied Chemistry ›› 1989, Vol. 0 ›› Issue (2): 56-59.
• Report • Previous Articles Next Articles
He Qijian, Meng Diya
Received:
Revised:
Published:
Online:
Abstract: For the lincar calibration of the monodisperse polymer species using more than two polydisperse standards with known number-average MW, an effective and fast one-variableapproach is deseribed. The equation ln M = A-BV containing only one variable B can besolved by the Newton's iteractive method, and the value of A is determined. The SEC(size exclusion chromatography) data of the characterized polyurethane samples were cor-rected for the chromatographic spreading and then used to demonstrate this computationalprocedure. The deviations of the number-average MW caculated by SEC from the resultsby osmotic msthed are acceptable.
Key words: Size exclusion chromatography, Calibration, Polydisperse polymer standards, Number-average MW
He Qijian, Meng Diya. CALIBRATION OF SEC COLUMN USNG MORE THAN TWO POLYDISPERSE POLYMER STANDARDS WITH KNOWN NUMBER-AVERAGE MW[J]. Chinese Journal of Applied Chemistry, 1989, 0(2): 56-59.