Chinese Journal of Applied Chemistry ›› 1989, Vol. 0 ›› Issue (2): 56-59.

• Report • Previous Articles     Next Articles

CALIBRATION OF SEC COLUMN USNG MORE THAN TWO POLYDISPERSE POLYMER STANDARDS WITH KNOWN NUMBER-AVERAGE MW

He Qijian, Meng Diya   

  1. Wuhan Chemical Industry Reseach Institute
  • Received:1988-05-26 Revised:1988-09-01 Published:1989-04-10 Online:1989-04-10

Abstract: For the lincar calibration of the monodisperse polymer species using more than two polydisperse standards with known number-average MW, an effective and fast one-variableapproach is deseribed. The equation ln M = A-BV containing only one variable B can besolved by the Newton's iteractive method, and the value of A is determined. The SEC(size exclusion chromatography) data of the characterized polyurethane samples were cor-rected for the chromatographic spreading and then used to demonstrate this computationalprocedure. The deviations of the number-average MW caculated by SEC from the resultsby osmotic msthed are acceptable.

Key words: Size exclusion chromatography, Calibration, Polydisperse polymer standards, Number-average MW