Abstract: Trace amount of iron in atmospheric particulate matter has been directly determined by graphite probe furnace atomic absorption spectrometry after collection with a graphite probe filter.The method was accurate,sensitive,and rapid.It has a recovery of 98% and a precision of 3.0% RSD, being checked against the certified values of NBS SRM-1648 urban particulate matter. The concentration of iron and absorbance (Peak-area mode) showed an excellent linear relation in the range of 0~100ng/ml and no memory effect.The characteristic mass and detection limit of iron were 14.6pg and 24.0pg, respectively.