All Solid-State Calcium Ion Selective Electrode Based on Carbon Nanotube/Ag/MoS2 Transducer

LUO Yunfenga,LIU Baoshuanga,LI Chunxiangb*()   

  1. aDrilling Technology Research Institute,Shengli Petroleum Engineering Co.,Ltd.,SINOPEC,Dongying,Shandong 257000,China
    bSchool of Chemistry and Chemical Engineering,Harbin Institute of Technology,Harbin 150001,China
  • Received:2018-09-20 Accepted:2018-12-05 Published:2019-06-01 Online:2019-06-03
  • Contact: LI Chunxiang
  • Supported by:
    Supported by the National Natural Science Foundation of Major Special Projects(No.2016ZX05022-006), the Innovation Talent project(Distinguished Young Scholars) of Harbin Science and Technology Bureau(No.2017RAYXJ015)

摘要:

构建了一种电势稳定性好的全固态钙离子选择电极(Ca2+-ISE),采用碳纳米管(CNT)/Ag/MoS2为转导层,自制的三脚架化合物作为钙离子载体,制备固体接触式Ca2+-ISE。 系统分析了全固态Ca2+-ISE的稳定性、能斯特斜率、响应范围、选择性系数等主要性能,发现制备的固体接触式Ca2+-ISE在钙离子浓度为1×10-6~1×10-1 mol/L范围内呈现线性能斯特响应,响应斜率为28.1 mV/decade。 CNT/Ag/MoS2的引入有利于提高固体接触式Ca2+-ISE的离子浓度线性响应范围,缩短电势平衡时间,降低测试能斯特斜率与理论值差,对于Ca2+-ISE的长期在线检测有重要研究意义。

关键词: 全固态电极, 钙离子选择性电极, 离子-电子转导层

Abstract:

An all-solid-state Ca2+ ion selective electrode(Ca2+-ISE) was prepared with carbon nanotube(CNT)/Ag/MoS2 as the ion-to-electron transducer and tripodal compound as the calcium ionophore. The stability, slope, response range, selectivity coefficient, etc. of the electrode were investigated. It is revealed that the electrode exhibits a 28.1 mV/decade Nernst slope in the range of 1×10-6~1×10-1 mol/L Ca2+ concentration. The transducer is helpful in enlarging the linear response range of the Ca2+ concentration, shortening the potential balance time and reducing the difference of the Nernst slope. All of these results are important for long-term online detection of calcium ions for the solid electrode.

Key words: all-solid-state electrode, calcium ion selectiveelectrode, ion-to-electron transducer