Chinese Journal of Applied Chemistry ›› 1995, Vol. 0 ›› Issue (4): 106-110.

• Full Papers • Previous Articles     Next Articles

Direct Determination of Lead and Antimony in Nickel Alloy by Platform Graphite Furnace Atomic Absorption Spectrometry

Xie Wenbing, Yao Jinyu, Hu Qinglan   

  1. Changchun Institute of Applied Chemistry, Chinese Academy of Sciences, Changchun 130022
  • Received:1994-09-21 Revised:1994-12-26 Published:1995-08-10 Online:1995-08-10

Abstract: The ashing temperature could be increased and some interferences from hy-drochloric acid,molybdenum and boron could be eliminated in title method by using Pd and Nias a matrix modifier.The method can be used directly for lead and antimony analysis inhigh temperature nickel-alloy without preseparation of the matrix. The characteristic mass is 28pg and 26pg, the detection limit is 0.13μg/g and 0.23μg/g,the recovery is 94%~103% and 95%~110%,the relative standard deviation is 6.4%~7.6% and 4.7%~7.5% for leadand antimony,respectively.

Key words: graphite furnace atomic absorption, nickel alloy, lead, antimony, matrix modifier